PPPL-4539
High Accuracy Wavelength Calibration For A Scanning Visible Spectrometer
Authors: Filippo Scotti and Ronald Bell
Abstract:
Spectroscopic applications for plasma velocity measurements often require wavelength accuracies ≤ 0.2Â. An automated calibration for a scanning spectrometer has been developed to achieve a high wavelength accuracy overr the visible spectrum, stable over time and environmental conditions, without the need to recalibrate after each grating movement. The method fits all relevant spectrometer paraameters using multiple calibration spectra. With a steping-motor controlled sine-drive, accuracies of ~0.025 Â have been demonstrated. With the addition of high resolution (0.075 aresec) optical encoder on the grading stage, greater precision (~0.005 Â) is possible, allowing absolute velocity measurements with ~0.3 km/s. This level of precision requires monitoring of atmospheric temperature and pressure and of grating bulk temperature to correct for changes in the refractive index of air and the groove density, respectively.
__________________________________________________
Submitted to: Review of Scientific Instruments (May 2010)
__________________________________________________
Download PPPL-4539 (pdf 201 KB 4 pp)
__________________________________________________