PPPL-4333

Thomson Scattering Density Calibration by Rayleigh and Rotational Raman Scattering on NSTX

Author: B.P. LeBlanc

Abstract:
The multi-point Thomson scattering (MPTS) diagnostic measures the profiles of the electron temperature Te(R) and density ne(R) on the horizontal midplane of NSTX. Normal operation makes use of Rayleigh scattering in nitrogen or argon to derive the density profile. While the Rayleigh scattering ne(R) calibration has been validated by comparison with other density measurements and through its correlation with plasma phenomena, it does require dedicated detectors at the laser wavelength in this filter polychromator based diagnostic. The presence of dust and/or stray laser light precludes routine use of these dedicated spectral channels for Thomson scattering measurement. Hence it is of interest to investigate the use of Raman scattering in nitrogen for the purpose of density calibration, since it could free up detection equipment, which could then be used for the instrumentation of additional radial channels. In this paper the viewing optics "geometrical factor" profiles obtained from Rayleigh and Raman scattering are compared. While both techniques agree nominally, residual effects on the order of 10% remain and will be discussed.
__________________________________________________

Published in: Review of Scientific Instruments 79, 10E737 (2008)

© (2008) American Institute of Physics.
This article may be downloaded for personal use only.
Any other use requires prior permission of the author
and the American Institute of Physics.


Presented at: 17th Topical Conference on High Temperature Plasma Diagnostics, May 11–15, 2008, Albuquerque, New Mexico.

__________________________________________________

Download PPPL-4333 (pdf 975.18KB 12pp)
__________________________________________________