Calibration Issues of the TFTR Multichannel Neutron Collimator
Authors: S. von Goeler, L.C. Johnson, M. Bitter, et al
Abstract: The calibration procedures for the detectors in the Neutron Collimator are reviewed. The absolute calibration was performed for the NE451 detectors, in situ, by moving a DT neutron generator in the TFTR vacuum vessel across each sight line. This calibration was transferred to other detectors in the same channel. Four new sight lines have been installed at a different toroidal location, which view the plasma through the vacuum vessel port cover rather than through thinned windows. The new detectors are cross-calibrated to the NE451 detectors with a "jog shot" procedure, where the plasma is quickly shifted in major radius over a distance of 30 cm. The jog shot procedure shows that scattered neutrons account approximately for 30% of the signal of the new central channels. The neutron source strength from the collimator agrees within 10 % with the source strength from global neutron monitors in the TFTR test cell. Detector non-linearity is discussed. Another special issue is the behavior of the detectors during T-puffs, where the DD/DT neutron ratio changes rapidly.